Pressure Sensor
(Order Code PS-DIN)
The
Pressure Sensor can be used to monitor pressure changes in gas-law experiments
in chemistry and physics, such as Boyle' s law (pressure vs. volume) and
Gay-Lussac' s law (pressure vs. absolute temperature). You can determine the
value of absolute zero on the Celsius temperature scale. The Pressure Sensor
can also be used to monitor reaction rates as a gas is produced in a chemical
reaction. Vapor pressure of various liquids and solutions can be monitored and
you can show the relationship between vapor pressure and absolute temperature.
Biology teachers can use the Pressure Sensor to monitor changes in the partial
pressure of oxygen or carbon dioxide gases in an enclosed atmosphere.
The Pressure Sensor can be used with any of the
following laboratory interfaces available from Vernier Software:
• MultiPurpose Lab Interface (MPLI)
• Voltage Input Unit (VIU)
• Universal Lab Interface
• Serial Box Interface
• Texas Instruments Calculator-Based Laboratory
(CBL) System
How the Pressure Sensor Works
The
heart of this circuit is the SenSym SCX1OOANC. It has a membrane which flexes
as pressure changes. This sensor is set up for absolute pressure measurement,
so one side of the membrane is a vacuum. The sensor produces an output voltage
which varies in a linear way with absolute pressure. It includes special
circuitry to minimize errors caused by changes in temperature. The 100 in the
part number indicates that the range of this sensor is 0 to 100 psi. The
"A" in the part number indicates that it measures absolute pressure.
Pressure Units Pressure can be measured in many different units.
We quote values here in several of the units shown below. Some equivalent
values for 1 atmosphere are:
1
atmosphere = 760 mm of Hg
= 101.325 kilopascals
= 29.92 in. of Hg (at 0°C)
= 1.013 bar
= 14.696 psi
We
provide an amplifier circuit that conditions the signal from the SCXI00 sensor.
With this circuit, the output voltage from the Pressure Sensor will be linear
with respect to pressure with 0.00 volts corresponding to a complete vacuum,
0.454 volts corresponding to 14.7 psi (one atmosphere), and 3.09 volts
corresponding to 100 psi. The output voltage increases 0.031 volts for each psi
increase. This allows the sensor to cover the range of 0 to 100 psi (6.8 atm).

There
is a plastic tube on the Pressure Sensor running from a port inside the box to
a three-way valve on the outside of the box. The two openings (or stems) on the
three-way valve have a small threaded end called a suer lock. You may attach
plastic or rubber tubing to one of the ports using the supplied adapter
(already mounted on extra tubing). The adapter accepts tubing with an inside
diameter of 0.125 inches (3.2 mm), and commonly available 3/16-inch plastic
tubing can be connected to it. You can also attach the 20-ml plastic syringe
included with the Pressure Sensor to this stem. The other port (pointing upward
in the figures below) of the three-way valve opens to the atmosphere and can
serve as a pressure release. As you set up your experiments, you can always
return the pressure to atmospheric pressure by opening this side stem. When the
blue control (or "off") handle is aligned with one of the stems, it
closes off this stem. Note: Since the side stem also has a luer lock, it is
possible to connect the syringe or the adapter with tubing in this side
position.

To
perform pressure-volume (Boyle's law) experiments, connect the plastic syringe
to one of the valve stems with a gentle 1/2
turn. Close the stem that leads into the
Pressure Sensor box using the blue control handle, and set the syringe to
any volume setting you like. Then close the open stem (see the figure below).
The pressure inside the syringe is now equal to atmospheric pressure at the volume
you selected. You are now set to collect pressure-volume data.

The
white, threaded adapter end of the long piece of plastic tubing can also be
connected to one of the valve stems with a full turn. The other end can be connected to an experimental apparatus. For
example, when doing a
pressure-temperature experiment (shown in the figure), the tubing can be
connected to a piece of glass tubing that is inserted into a one-hole rubber
stopper. This stopper can in turn be
inserted into an Erlenmeyer flask to provide a constant-volume gas sample.
Pressure-temperature data (Gay-Lussac's law) can be collected using a
temperature probe or thermometer, along with the Pressure Sensor.

The
SCX 100ANC pressure sensor is fairly durable, but it is designed only for use
with non corrosive gases such as air, helium, nitrogen, etc. Do not get it wet.
Vernier
Software also manufactures a Barometer (order code BAR-DIN) that covers a
pressure range of 24 to 32 inches of mercury (0.8 to 1.05 atmospheres) absolute
pressure. It is designed primarily for monitoring atmospheric pressure as part
of weather studies. It has a much higher sensitivity over its limited pressure
range.
Calibration
There
are several ways to handle calibration with this sensor. In some cases, the
circuit can be used without calibration. Often you are interested in only
relative pressures, and you can use the raw voltage output of the Pressure
Sensor. With the Pressure Sensor as shipped, voltage is proportional to
pressure. In most cases, you can simply load the appropriate calibration file
that was included on your laboratory interfacing program from Vernier Software.
This calibration will be satisfactory for most experiments using the Pressure
Sensor.
For
the very best results, you can calibrate this unit yourself using one of the
methods described below. You will need to know the local air pressure, and/or
have a reliable pressure measurement instrument.
Two Point Calibration
This
is the calibration procedure we use with all of our sensors. You simply do a
two-point calibration as described in the program manual using two different
pressures, both measured with another pressure gauge. One pressure is easy to
get in this case— simply use atmospheric pressure as measured by a barometer.
For the other pressure, you can:
•
Use the syringe provided with the Pressure Sensor to produce a pressure very
near zero. We have found that this method works very well. Connect the syringe
to the Pressure Sensor and open the unused stem of the three-way valve by
aligning the blue valve control with the stem that leads into the Pressure
Sensor box. Push the plunger on the syringe all the way in and close the unused
stem leading to the atmosphere by aligning the blue valve control with it. To
produce near zero pressure, pull the plunger out to the 20-cc position. If your
syringe and valve have a tight seal, the pressure will be <0.01 atm, and you
can call it zero for calibration.
•
Apply pressure with a pump, measuring it at the same time with a pressure
gauge.
Whatever
method you use, the input should be named "Pressure," and appropriate
units should be entered as part of the calibration.
One-Point Calibration
This
calibration method uses the predetermined sensitivity of the Pressure Sensor to
simulate a two-point calibration. It cannot he used if you have changed the
gain potentiometer inside the box. Follow the Instructions on the screen and in
your program manual to begin calibration. The input should be named
"Pressure" and appropriate units should be entered. For the first of
the two calibration points, open the unused stem of the three-way valve to
expose the Pressure Sensor port to atmospheric pressure. Do this by aligning
the blue valve control with the stem leading to the inside of the Pressure
Sensor. Read a laboratory barometer and enter its pressure reading as the known
pressure. Also, note the voltage reading.
For
the second point, close the unused stem by aligning the blue valve control with
it. Blow on the tubing connected to the three-way valve of the Pressure Sensor,
so that the voltage reported increases. Now press <Return>, while noting
the voltage. Take this as the second calibration point and calculate the change
in voltage between the two calibration points. Calculate what the pressure was
using the known sensitivity of the Pressure Sensor. Use the equation:
Pressure
(in. psi) = first pressure reading + voltage change /0.031 volts/psi.
If
you use different pressure units, make the appropriate conversion. This
calculation gives you the pressure at the second calibration point. It uses the
known sensitivity of the unit.
Adjusting the Pressure Sensor
There
is an offset potentiometer that you can use to increase or decrease the output
signal. This potentiometer can be adjusted without opening the box. Place a
small jeweler's screwdriver through the hole in the end of the box and you will
notice a slotted screw that can be turned. This is a l 5-turn potentiometer, so
the adjustment can be made very gradually. One way to adjust the calibration of
the Pressure Sensor is to adjust this potentiometer. There is also a gain
control potentiometer inside the box. Adjusting this potentiometer requires
opening the box, and is not recommended for normal use. When this unit is
shipped, we set the gain potentiometer in the fully clockwise position. This
sets the gain at its lowest value. This potentiometer can be turned
counter-clockwise to increase the sensitivity. The maximum sensitivity is about
290 mV per psi. If you adjust the potentiometers, test the response of the unit
by attaching a plastic tube to the input port, and blow and suck on it. Make
sure that the voltage increases when you blow on the tube and decreases when
you suck on the tube.
A
complete schematic of the Pressure Sensor is available. Contact Vernier
Software.
Specifications
•
Sensing element: SenSym SCX100ANC
•
Pressure range (as shipped): 0 to 6.8 atm (0 to 100 psi)
•
Maximum pressure that the sensor can tolerate without permanent damage:
10.2
atm or 150 psi
•
Sensitivity (as shipped, minimum) 454 mV/atm or 31 mV /psi
(with
gain adjusted to maximum) 4.2 V/atm or 290 mV /psi
•
Resolution (as shipped) using a 12-bit, 5 volt A/D converter (ULI II, Serial
Box):
0.00275
atm
•
Resolution (as shipped)using a 10-bit, 5 volt A/D converter (CBL, original
ULI):
0.011
atm
•
Combined linearity and hysteresis: typical to 1% full scale, maximum to 0.5%
full scale
•
Response time: 100 microseconds
Suggested Experiments
Boyle's Law (Pressure vs.
Volume)
Boyle's
law is a classic physics/chemistry concept that can be easily demonstrated
using the Pressure Sensor. One easy way to do this is to use the plastic
syringe included with the Pressure Sensor. Connect the plastic syringe to one
of the valve stems with a gentle 1/2
turn. Close the stem that leads into the Pressure Sensor box using the blue
control handle and set the syringe to any volume setting you like. Then close
the open stem (see the figure on page 2). The pressure inside the syringe is
now equal to atmospheric pressure at the volume you selected. You are now set
to collect pressure-volume data. Take data as you change the volume. The
syringe is marked in volume units (cc or ml). You can both increase and
decrease the volume. Sample data collected with this sensor and the syringe are
graphed below:

Sample
Boyle's Law Graphs
Gay-Lussac's Law (Pressure
vs. Absolute Temperature)
Gay-Lussac's
law states that if the volume remains constant, the pressure of a container of
a gas is directly proportional to its absolute temperature. The white, threaded
adapter end of the long piece of plastic
tubing can be connected to one of the valve stems with a full turn. The other
end can be connected to a piece of glass tubing that is inserted into a
one-hole rubber stopper. This stopper can in turn be inserted into an
Erlenmeyer flask to provide a constant-volume gas sample. Pressure-temperature
data can be collected using a temperature probe or thermometer, along with the
Pressure Sensor. Place the flask in water baths of different temperatures. Take
data on how the pressure changes with temperature changes. remember that all
temperatures should be measured using the Kelvin temperature scale. Using the
same apparatus, pressure and temperature data may be extrapolated to determine
a Celsius temperature value for absolute zero.
Studying Chemical Reactions
by Monitoring Pressure
Many
chemical reactions produce gases that can cause a pressure increase in a sealed
container. The pressure change can be used as a way of monitoring the rate of
reaction. You can change concentrations or temperatures, or add catalysts, to
see how the reaction rate changes. As a very simple example, the graph below
was made using the Pressure Sensor to monitor {he reaction of Alka-Seltzers and
water. We made a hole in the top of a plastic 35-mm film container. We then ran
a tube from the Pressure Sensor through the hole and sealed it with silicone
sealant. We then placed a piece of an Alka-Seltzer tablet and 10 ml of water in
the container and sealed it. The graph below shows the pressure increase as the
reaction takes place. Note the sudden drop in pressure when the lid of the
container pops off.

Using
the Pressure Sensor to Monitor the Rate of Reaction
Pressure in Liquids
If you measure (he pressure al the end of a
long plastic tube forced underwater, you can indirectly measure depth. Connect
the tubing to the input port of the Pressure Sensor and then put the end of the
tube under water. The pressure reading will increase 0.0965 atmospheres for
every meter below the surface of the water. If you use different pressure
units, you will need so make the appropriate conversion.
Note:
if you measure depth in this way, the depth you are measuring is to the top of
the air, which extends up the tube for a short distance If this measurement
error bothers you, you can simply calibrate your depth measurement system when
the end of the tube is al known depths and automatically correct for this.
Additional
experiments are available in our Chemistry
with Computers or Chemistry with CBL books.
Contact Vernier Software for details.
Flinn QuickStartsTM
Pressure-Pressure Probe ()
The Connections:
1. Connect the
Pressure Probe to Port 1 of the Serial Box Interface.
2. Connect the Serial
Box Interface to the COM I or COM2 port of the computer using the cable
provided. (Use the 9-pin to 25-pin adapter if necessary.)
3. Plug the Serial
Box Interface into an electrical outlet using the AC adapter.
The Set-up/Customizing Data Logger:
1. Open Data Logger program; Click on Start, Program, Vernier, Logger Pro.
Note: If a message
box reading “Warning! Cannot find the interface box” appears:
a) Check all
connections,
b) Determine which
port (COM1 or COM2) is accepting the cable from the Serial Box
Interface, and
c) Click on the
appropriate button (“COM1” or “COM2”).
A default graph call
untitled- 1” will appear.
The right mouse button Is used only to access Data logger
menus. Use the left mouse button for all other operations.
2.
Under the “File” menu, choose “Open... ”. A new window will appear,
listing the various experiment files for Data Logger.
3.
From the list, double-click on “Pressure-pressure Probes” choose “Pressure Probe” to open the file for the direct temperature
measurements. You may need to scroll the list by clicking on the left mouse
button and dragging the cursor across the list to find these files.
5.
The “Setup” menu can be used to
specify
Sensor [Sensor setup, Calibration, and Details(units can be specified
here)]
Data Collection [Mode, Sampling (Rate)]
and
Interface details (Com port).
6. You do not need to
customize the axes for the introductory activity. However, if you would like to
customize the axis:
a) Double-click
anywhere on the graph. This cases the “Axes” dialog box to appear.
b) Click and hold the
cursor on the y-axis label (which currently reads
“Temp 1 & Temp
2”). Move the cursor to “Temperature 1” to collect data using Port 1 only. If
you are using temperature probes in both ports, you do not have to do this.
c) To change the
y-axis temperature range, simply Double-click inside the upper and/or lower
limit value box and enter the desired value.
d) The x-axis should
already be in units of “Time.” You may change the values in the upper and/or
lower limit value boxes in the same manner as you did for the y-axis above.
You are now ready to
collect data. Simply click
“Collect” when you are ready to begin;
click “Stop” to discontinue data collection.
Introductory Activity:
Calibration:
1. Open the “Pressure
probe” file and load the calibration file, as described above.
2. Evacuate system
with isoteniscope connected determine pressure inside system.
3. Place the Standard
Temperature Probe in the water bath.
4. Click on
”Collect”.
5. Click “Stop” to
discontinue data collection.
6. Heat a water bath
to boiling!
7. Place the Standard
Temperature Probe in the water bath.
8. Click on
”Collect”.
9. Use Calibration in
the “Experiment” menu to calibrate the sensor using the water baths at 0 °C and 100 °C.